NATA-accredited calibration is done by a laboratory that NATA has assessed and accredited for that type of measurement. NATA-traceable calibration is done against reference equipment that was itself calibrated by a NATA-accredited laboratory. Both trace back to national measurement standards. The difference is who is accredited.
What “traceable” means
Every measurement needs something to be checked against. Your torque wrench is checked against a reference torque device. That reference was checked against a more accurate one, and so on, up to Australia’s national measurement standards. That unbroken chain is traceability.
When calibration is NATA-traceable, the step above yours in the chain was done by a NATA-accredited laboratory. Your report shows your instrument’s results against that traceable reference.
What “accredited” means
NATA accredits laboratories to ISO/IEC 17025. An accredited laboratory has been assessed for its methods, people, equipment and how it works out measurement uncertainty, for a defined scope of measurements. Its reports can carry the NATA logo for work within that scope.
Which one do you need?
It depends on the wording your customer, auditor or quality system uses.
- Many quality systems, including ISO 9001, ISO 3834 and AS/NZS 5131, generally need equipment calibrated against traceable standards. In many cases, a NATA-traceable report is enough.
- Some contracts or regulators specifically ask for calibration by a NATA-accredited laboratory, or a report with the NATA logo. Then traceable isn’t enough.
If the wording says “calibrated”, “traceable” or “traceable to national standards”, a traceable report usually fits. If it says “NATA-accredited”, “NATA-endorsed” or asks for the NATA logo, you need an accredited laboratory.
How to check before you book
- Find the exact wording in your contract, specification or quality manual.
- Send it to your calibration provider before booking.
- Ask what the report will show: the instrument, test points, readings, reference equipment and date.
Where PCC fits
All PCC calibrations are NATA-traceable. We calibrate your equipment against reference equipment calibrated by NATA-accredited laboratories. PCC is not a NATA-accredited laboratory. If your customer needs accredited calibration, tell us before booking and we’ll confirm the options for your equipment.